EUREKA™-IP
&
EUERKA™-LF
INFUSION PUMP
OPERATOR’S MANUAL
Universal Medical Technologies Inc.
Page view 0
1 2 3 4 5 6 ... 23 24

Summary of Contents

Page 1

EUREKA™-IP&EUERKA™-LFINFUSION PUMPOPERATOR’S MANUALUniversal Medical Technologies Inc.

Page 2 - ™ Infusion Pump, Please call

Utilize Eureka supplied Power Adapter ONLY. Recharge time is between 30-60 minutes during non-operation. Charging during use may result in slightly l

Page 3 - Contents

3.4 BATTERY LIFE (Typical Charge Times)Initial charge time (average) 22 minutesSubsequent charge times (range) 30 minutes- 60 minutes“Conditioned Batt

Page 4

3.6 PLACING BAG IN PUMPPosition Bag in Pump as follows:Place pump on flat surface. Place bag so that SpikeFlange rests in slot before case outlet, an

Page 5

3.7 PUMP OPERATIONS3.7.1 USING THE EUREKA ™ INFUSION PUMP1. Engage EUREKA™ administration set and IV solution container per administration Set instru

Page 6

3.7.2 WARNINGS AND ALARMS1. POWER Light GREEN (2) indicates pump is ON when illuminated. POWER up sequence atpowering ON of unit consists of a conti

Page 7 - D5W vs NS @ 100 ml/hr

FUNCTION KEYSSTART-Starts or restarts infusion cycle- Press once when Power is ONSTOP- hold press (2 seconds or 2 beeps) to stop or pause operationPOW

Page 8 - Pressure Effect (NS)

TYPICAL FLOW RATE RESULTS FOR ADMINISTRATION SETLabeled Set Rate 100 ml/hrMax. fill capacity 110 mlsRetained Volume 2-4 mlsTubing

Page 9

DisposableDelivery accuracy : +/- 12% based on 100 ccs Normal Saline Viaflex containerFluid pathway is sterile and nonpyrogenic. Priming Volume is 1.2

Page 10

To Disconnect:1. At pump signal of Check Status verify complete infusion.a. If infusion has not completed then check the IV tubingline for crimping ,

Page 11

4.1 GLOSSARYAdministration Set- Tubing and spike attachment that connects solution container and patient’s access deviceAspirate - To allow air to ven

Page 12

This guide applies to the EUREKA-IP™ INFUSION PUMP Catalog# 11016© 2001 Universal Medical Technologies Inc.All rights reservedPrinted in USAUniversal

Page 13 - 3.7 PUMP OPERATIONS

4.2 PUMP CLEANING & MAINTENANCEUse soft damp cloth only.Cloth may be dampened with one of the following solutionsSoap and water70% Ethyl Alcohol70

Page 14

4.4 SPECIFICATIONSModel - IP Catalog Number 11016 Eureka-IP Pump with Accessories (US)Model - LF Catalog Number 11017 Eureka-LF Pump with Accessories

Page 15

Universal Medical Technologies (UMT) warrants to the original purchaser that the Infusion Pump (not including accessories) is free from defects inmat

Page 16 - Flow Profiles

5.0 PRODUCT TRACKING FORM23

Page 18 - (Feburary 2003)

Contents1.0 INTRODUCTION ... 42.0 GENE

Page 19 - 4.0 REFERENCE SECTION

2.1.2 PUMP PACKAGINGTHE EUREKA ™ INFUSION PUMPS are sold as follows1- EUREKA ™ INFUSION PUMP (IP or LF)1- Operator Manual1-Battery Recharger / Power

Page 20

52.2 IV SOLUTION CONTAINERSThe EUREKA ™ INFUSION PUMPS will accommodateBaxter 50 ml and 100 ml IV solution containers.Abbott 50 ml and 100 ml IV solu

Page 21 - ™ Infusion Pump

62.3.1 AVAILABLE FLOW RATESCurrently the Eureka-IP Administration Set flow rates are: 50 ml/hr EUREKA ™ IV ADMINISTRATION SET100 ml/hr EUREKA ™ IV

Page 22

2.3.3 FLOW RATE VARIABLESTime (Minutes)020406080100120140Average Flow Rate47F77FNormal Saline 100 ml (Baxter)100 cc/hrInitial Temp: 47F vs 77FEffect o

Page 23 - 5.0 PRODUCT TRACKING FORM

Effect of Pressure (NS=Normal Saline)159131721252933374145495357616569737781858993Minutes60708090100110120130cc/hr (cumlative).65 psi0 psiPressure Eff

Page 24

3.0 OPERATING INSTRUCTIONS3.1 CHANGING BATTERIES / RECHARGING / DC ADAPTERWhen battery door is down., disconnect battery ca-bling. Slide battery pac