EUREKA™-IP&EUERKA™-LFINFUSION PUMPOPERATOR’S MANUALUniversal Medical Technologies Inc.
Utilize Eureka supplied Power Adapter ONLY. Recharge time is between 30-60 minutes during non-operation. Charging during use may result in slightly l
3.4 BATTERY LIFE (Typical Charge Times)Initial charge time (average) 22 minutesSubsequent charge times (range) 30 minutes- 60 minutes“Conditioned Batt
3.6 PLACING BAG IN PUMPPosition Bag in Pump as follows:Place pump on flat surface. Place bag so that SpikeFlange rests in slot before case outlet, an
3.7 PUMP OPERATIONS3.7.1 USING THE EUREKA ™ INFUSION PUMP1. Engage EUREKA™ administration set and IV solution container per administration Set instru
3.7.2 WARNINGS AND ALARMS1. POWER Light GREEN (2) indicates pump is ON when illuminated. POWER up sequence atpowering ON of unit consists of a conti
FUNCTION KEYSSTART-Starts or restarts infusion cycle- Press once when Power is ONSTOP- hold press (2 seconds or 2 beeps) to stop or pause operationPOW
TYPICAL FLOW RATE RESULTS FOR ADMINISTRATION SETLabeled Set Rate 100 ml/hrMax. fill capacity 110 mlsRetained Volume 2-4 mlsTubing
DisposableDelivery accuracy : +/- 12% based on 100 ccs Normal Saline Viaflex containerFluid pathway is sterile and nonpyrogenic. Priming Volume is 1.2
To Disconnect:1. At pump signal of Check Status verify complete infusion.a. If infusion has not completed then check the IV tubingline for crimping ,
4.1 GLOSSARYAdministration Set- Tubing and spike attachment that connects solution container and patient’s access deviceAspirate - To allow air to ven
This guide applies to the EUREKA-IP™ INFUSION PUMP Catalog# 11016© 2001 Universal Medical Technologies Inc.All rights reservedPrinted in USAUniversal
4.2 PUMP CLEANING & MAINTENANCEUse soft damp cloth only.Cloth may be dampened with one of the following solutionsSoap and water70% Ethyl Alcohol70
4.4 SPECIFICATIONSModel - IP Catalog Number 11016 Eureka-IP Pump with Accessories (US)Model - LF Catalog Number 11017 Eureka-LF Pump with Accessories
Universal Medical Technologies (UMT) warrants to the original purchaser that the Infusion Pump (not including accessories) is free from defects inmat
5.0 PRODUCT TRACKING FORM23
Contents1.0 INTRODUCTION ... 42.0 GENE
2.1.2 PUMP PACKAGINGTHE EUREKA ™ INFUSION PUMPS are sold as follows1- EUREKA ™ INFUSION PUMP (IP or LF)1- Operator Manual1-Battery Recharger / Power
52.2 IV SOLUTION CONTAINERSThe EUREKA ™ INFUSION PUMPS will accommodateBaxter 50 ml and 100 ml IV solution containers.Abbott 50 ml and 100 ml IV solu
62.3.1 AVAILABLE FLOW RATESCurrently the Eureka-IP Administration Set flow rates are: 50 ml/hr EUREKA ™ IV ADMINISTRATION SET100 ml/hr EUREKA ™ IV
2.3.3 FLOW RATE VARIABLESTime (Minutes)020406080100120140Average Flow Rate47F77FNormal Saline 100 ml (Baxter)100 cc/hrInitial Temp: 47F vs 77FEffect o
Effect of Pressure (NS=Normal Saline)159131721252933374145495357616569737781858993Minutes60708090100110120130cc/hr (cumlative).65 psi0 psiPressure Eff
3.0 OPERATING INSTRUCTIONS3.1 CHANGING BATTERIES / RECHARGING / DC ADAPTERWhen battery door is down., disconnect battery ca-bling. Slide battery pac